Precision LCR Meter LCR-8000G Series USER MANUAL GW INSTEK PART NO. 82CR-81010MD1 ISO-9001 CERTIFIED MANUFACTURER This manual contains proprie
GETTING STARTED 19191919 Rear Panel Overview GPIBRS232WARNINGSER.NO. LABELAC 115 / 230V 50 / 60 HzFUSE 250V T 3A115VGPIBPortRS-232CPortD
GETTING STARTED 21212121 Tilt Stand & Power Up Tilt stand Low angle High angle LCR-8000G Series User Manual 22222222 Power up Pan
GETTING STARTED 23232323 Select AC mains frequency (50/60Hz) Background Although the LCR-8000G works under both 50 and 60Hz power frequencies,
GETTING STARTED 25252525 Fixture connection Panel operation 1. Discharge the test component before connecting the fixture set. 2. Connect e
GETTING STARTED 27272727 Optional settings To hide the drive Voltage/Current, press the Code key, type 80, then press Enter. Set the Range (int
GETTING STARTED 29292929 10b. Select Repetitive measurement Press the Sing/Rep key to select Repetitive (automatic trigger) measurement. Press
GETTING STARTED 31313131 8. Enter Run menu Press F6 (Run). The Run menu opens. Page80 9. Set Single or Repetitive Press the Sing/Rep key to se
GETTING STARTED 33333333 8b. Set vertical scale (Absolute + Manual fit) Press F2 (Abs/%) to select Abs, then press F3 (Manual/Auto fit) to sele
GETTING STARTED 35353535 Wire capacitance When measuring the wire capacitance, the fixture clips that are marked with HF(High Force)/HS (High S
BASIC MEASUREMENT 37373737 Measurement Item Description In general, two items, primary and secondary, are combined in a single measurement. The
TABLE OF CONTENTS 3333 Table of Contents SSSSAFETY INSTRUCTIONAFETY INSTRUCTIONAFETY INSTRUCTIONAFETY INSTRUCTION ...
BASIC MEASUREMENT 39393939 Series formula +=211QLLPS Q=quality factor Parallel formula +=211QLLSP Q=quality factor When to
BASIC MEASUREMENT 41414141 Capacitance (C) Background Capacitance measures the amount of electronic charge stored between two terminals. Displ
BASIC MEASUREMENT 43434343 Reactance (X) and Susceptance (B = 1/X) Background Reactance measures the imaginary part of Impedance (Z) caused by
BASIC MEASUREMENT 45454545 Quality factor (Q) and Dissipation factor (D) Background Both Quality factor and its reciprocal, Dissipation factor
BASIC MEASUREMENT 47474747 Measurement Mode Overview Enter measurement mode Type AC C, L, X, B, Z, Y, Q, D, R, G, θ DC Rdc Panel operation
BASIC MEASUREMENT 49494949 Delta mode (Pass/Fail test) For Pass/Fail test details, see page58. Show circuit model or scale (pass/fail) Backg
BASIC MEASUREMENT 51515151 Set measurement range to Auto Background The measurement range is an internal parameter defining the search range f
BASIC MEASUREMENT 53535353 Select frequency step resolution For frequency increase/decrease using Up/Down keys, fine and coarse step settings a
BASIC MEASUREMENT 55555555 Running Measurement Select Single measurement Background The data capture can be manually controlled (Single) or au
BASIC MEASUREMENT 57575757 DC AC≤ 100Hz AC≤ 2kHz AC> 2kHz AC≥ 1MHz Slow 900ms 1.3s 600ms 600ms 620ms Med 120ms 1.2s 470ms 450ms 47
TABLE OF CONTENTS 5555 SAFETY INSTRUCTION This chapter contains important safety instructions that you must follow when operating or storing an LC
PASS-FAIL MODE 59595959 Multi-step configuration Overview ...70 Configure be
PASS-FAIL MODE 61616161 RP Parallel resistance θ Angle For detailed description of each item, see page37. Configure beep setting Background
PASS-FAIL MODE 63636363 Select test item and scale (Pass/Fail test) Test item To select the first measurement item, press F1 repeatedly. C L X
PASS-FAIL MODE 65656565 Frequency (except Rdc) Press the Left/Right key repeatedly to move the cursor to Frequency, and use the numerical and u
PASS-FAIL MODE 67676767 Result > Hi HI Result < Lo LO Lo < Result < Hi (Pass) PASS Run in Percentage mode 1. Press F
PASS-FAIL MODE 69696969 Backspace All clear Clear Hi and Lo values are automatically swapped if necessary Hi and Lo Swapped 4. The
PASS-FAIL MODE 71717171 Parameters No. of step Maximum 30 for each program No. of program Maximum 64 Drive Voltage (1mV step) 10mV ~ 2V (DC
PASS-FAIL MODE 73737373 Configure the Average Background The Average function sets the number of samples used, which are then averaged as the
PASS-FAIL MODE 75757575 Move cursor (Left/Right keys) JKLMNOPQRS Enter the letter (Down key) program name: N_ Delete one letter (
PASS-FAIL MODE 77777777 Set frequency Move the cursor to Freq column. Enter the frequency using the numerical keys and unit keys. FuncFreq500.0
SAFETY INSTRUCTION 7777 Fuse WARNING • Fuse type: T3A/250V • Make sure the correct type of fuse is installed before power up. • To ensure fir
PASS-FAIL MODE 79797979 Copy (duplicate) program step Background Copying the step inserts a new, identical step next to the current step (= th
PASS-FAIL MODE 81818181 3. In Manual (single) mode press F1 (Start) or the Trigger key to manually start the program. The test results show u
PASS-FAIL MODE 83838383 Multi-Step Program File Operation Save program Save (overwrite) Press F4 (Save) to save the program being edited. A co
PASS-FAIL MODE 85858585 Recall (load) existing program Panel operation 1. Press F5 (File). The file menu appears. F 5 2. Press F1 (Load). T
GRAPH MODE 87878787 GRAPH MODE The graph function shows the component characteristics in visual manner. Voltage and Frequency sweep are selecta
GRAPH MODE 89898989 Select measurement item Range Ls Series inductance θ Angle Lp Parallel inductance Rs Series resistance Q Quality fact
GRAPH MODE 91919191 Unit Mismatched If a value outside of the range is entered, the closest available value is automatically selected.
GRAPH MODE 93939393 Unit Mismatched If a value outside of the range is entered, the closest available value is automatically selected.
GRAPH MODE 95959595 Backspace All clear Clear If a wrong unit is entered, the value is cancelled. Unit Mismatched If a value outs
GRAPH MODE 97979797 If a wrong unit is entered, the value is cancelled. Unit Mismatched If the entered Hi level is lower than the Lo l
SAFETY INSTRUCTION 9999 Power cord for the United Kingdom When using an LCR-8000G series LCR meter in the United Kingdom, make sure the power c
GRAPH MODE 99999999 Set vertical axis (Auto + Percentage) Background The Y (vertical) axis configuration is available for: • Manual or Automat
GRAPH MODE 101101101101 Speed / Step Setting Select measurement speed (capture timing) Background The speed is the same setting used in the ba
GRAPH MODE 103103103103 Running Graph Measurement Run measurement Panel operation 1. When the configuration is completed, press F4 (Start) to
GRAPH MODE 105105105105 Adjust vertical scale Background When the measured data does not fit into the original vertical scale, use this funct
REMOTE CONTROL 107107107107 REMOTE CONTROL This chapter describes basic aspects of IEEE488.2 based remote control. Both RS-232C and GPIB interf
REMOTE CONTROL 109109109109 Configure GPIB interface Connection Connect the GPIB cable to the rear panel port: 24-pin female connector. Pin a
REMOTE CONTROL 111111111111 Command Syntax Compatible standard • IEEE488.2, 1992 (fully compatible) • SCPI, 1994 (partially compatible) Comman
REMOTE CONTROL 113113113113 *stb? Returns the current contents of the Status Byte with the Master Summary bits as an integer in the range 0 to
REMOTE CONTROL 115115115115 :meas:speed <disc> :meas:speed? Selects or returns measurement speed. Send parameter: max, fast, med, slow Re
REMOTE CONTROL 117117117117 :multi:new <filename> Create a new multi-step program. Example: :multi:new demo (file name demo) :multi:res?
GETTING STARTED 11111111 Fixture connection Fixture structure...24 Fixture connection .
REMOTE CONTROL 119119119119 :graph:freq? Returns the frequency if the sweep mode is drive level :graph:lev <NR3> Set the drive level if t
REMOTE CONTROL 121121121121 :graph:mk? Returns the measurement from the current marker position. Returned parameter: Depending on the measured
CALIBRATION 123123123123 CALIBRATION Overview Background Calibration (trimming) eliminates stray capacitance and series impedance from the tes
CALIBRATION 125125125125 Trim the LCR-8000G Fixture setting Prepare the fixture accordingly. (In order to run a complete trimming, both O/C an
CALIBRATION 127127127127 5. The trimming starts and ends automatically. SHORT CIRCUIT TRIMIN PROGRESSALL FREQUENCY Trim pass The displ
FAQ 129129129129 FAQ Q1. The beep keeps sounding. A1. The beep sounds according to the pass/fail test result, which in this case is set to rep
APPENDIX 131131131131 O/S Trimming Points All-Frequency O/S Trimming Point List The frequency of the extra trimming point is the measurement f
APPENDIX 133133133133 ≦10 kHz O/S Trimming Point List The frequency of the extra trimming point is the measurement frequency you were using be
APPENDIX 135135135135 |Z| vs L, C Chart Conversion method: CF→|Z| CFreq.Z LF→|Z| LCR-8000G Series User Manual 136136136136 Accuracy Def
APPENDIX 137137137137 Specification Test Frequency LCR-8101G DC, AC:20Hz~1MHz LCR-8105G DC, AC:20Hz~5MHz LCR-8110G DC, AC:20Hz~10MHz
GETTING STARTED 13131313 Package Contents Ensure all the package contents are included and defect-free before using the LCR-8000G. If any of th
APPENDIX 139139139139 AC Drive Signal Frequency Accuracy 5 Digits, ±0.005% LCD Display 320×240 Dot Matrix Interface RS-232, GPIB GPIB Addr
APPENDIX 141141141141 Declaration of Conformity We GOOD WILL INSTRUMENT CO., LTD. (1) No. 7-1, Jhongsing Rd., Tucheng City, Taipei County, Tai
INDEX 143143143143 graph mode... 87 configuration...88 run ...
GETTING STARTED 15151515 Model Comparison Major Model Differences Model LCR-8101G LCR-8105G LCR-8110G Measurement Frequency 20Hz~1MHz 20Hz~
GETTING STARTED 17171717 Calibration key Calibration Enters the calibration mode. See page123 for calibration details. Unit keys Enters unit wh