Gw-instek GDS-800 Series Programming Manual Bedienungsanleitung Seite 11

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 89
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 10
GDS-806/810/820/840 Programming Manual
10
The RS232 connection testing
If you want to test whether the RS-232 connection is working or not, you can send a
command from computer. For instance, using a terminal program send the query
command
*idn?
should return the Manufacturer, model number, serial number and firmware version
in the following format:
GW,GDS-820,P920130,V.2.01
If you do not receive a proper response from the oscilloscope, please check if the
power is on, the RS-232 configurations are the same on both sides, and all cable
connections are active.
Seitenansicht 10
1 2 ... 6 7 8 9 10 11 12 13 14 15 16 ... 88 89

Kommentare zu diesen Handbüchern

Keine Kommentare