COMMAND DETAILS
83
Return parameter
MEAS<X>
The automatic measurement number
from 1 to 4.
Note
The measurement source(s), measurement number
and measurement type must first be set before a
measurement result can be returned.
Example
:MEASUrement:MEAS1:SOUrce1 CH1
:MEASUrement:MEAS1:TYPe PK2PK
:MEASUrement:MEAS1:VALue?
5.000E+0
Selects channel 1 as the source for measurement 1
and sets measurement 1 to peak to peak
measurement. The result returns the peak to peak
measurement.
:MEASUrement:MEAS<X>:MAXimum
Description
Returns the maximum measurement results for the
selected measurement from the last time the
statistics were reset. This is a statistics related
command.
Syntax
:MEASUrement:MEAS<X>:MAXimum?
Related
Commands
:MEASUrement:STATIstics:MODe
Parameter
MEAS<X>
The automatic measurement number
from 1 to 4.
Example
:MEASUrement:MEAS3:SOUrce1 CH1
:MEASUrement:MEAS3:TYPe PK2PK
:MEASUrement:STATIstics:MODe ON
:MEASUrement:MEAS3:MAXimum?
2.800E-02
Returns the maximum measurement result for
measurement number 3.