Gw-instek GDS-3000 Series Programming Manual Bedienungsanleitung Seite 61

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 103
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 60
COMMAND DETAILS
61
:MEASure:FREQuency
Query
Description
Returns the frequency value.
Syntax
:MEASure:FREQuency{?}
Return parameter
Returns the frequency as <NR3>.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:FREQuency?
1.0E+3
Selects Channel 1, and then measures the
frequency.
:MEASure:NWIDth
Query
Description
Returns the first negative pulse width timing.
Syntax
:MEASure:NWIDth{?}
Return parameter
Returns the negative pulse width as <NR3>
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH 1
:MEASure:NWIDth?
Selects Channel 1, and then measures the negative
pulse width.
:MEASure:PDUTy
Query
Description
Returns the positive duty cycle ratio as percentage.
Syntax
:MEASure:PDUTy{?}
Return parameter
<NR3>