REMOTE CONTROL
101
Returns the bit sum of the Standard Event
Status Enable register.
Queries the Standard Event Status (Event) register.
The Event Status register is cleared after it is read.
Returns the bit sum of the Standard Event
Status (Event) register and clears the register.
Queries the manufacturer, model name, serial
number, and firmware version of the PSU.
Returns the instrument identification as a
string in the following format:
GW-INSTEK,PSU-2076,TW123456,01.00.20110101
Manufacturer: GW-INSTEK
Model number : PSU-3036
Serial number : TW123456
Firmware version : 01.00.20110101
The *OPC command sets the OPC bit (bit0) of the
Standard Event Status Register when all current
commands have been processed.
The *OPC? Query returns 1 when all the
outstanding commands have completed.