PSU Programming Manual
84
[:SOURce]:VOLTage[0]:LIMit:AUTO
Sets whether to limit the voltage setting so that it
does not exceed the OVP setting or become lower
than the UVP setting.
If you enable the limit when the OVP setting is
lower than the voltage setting, the OVP setting will
be set to 105 % of the voltage setting.
If you enable the limit when the UVP setting is
higher than the voltage setting, the UVP setting
will be set equal to the voltage setting.
[:SOURce]:VOLTage[0]:LIMit:AUTO {<bool>|OFF|ON}
[:SOURce]:VOLTage[0]:LIMit:AUTO?
Disable the limit setting
Returns the setting in <bool> format.
SOUR:VOLT:LIM:AUTO 0
Disables the limit setting.
[:SOURce]:VOLTage[0]:LIMit:LOW
Sets or queries the under voltage (UVL) trip point.
[:SOURce]:VOLTage[0]:LIMit:LOW
<NR2>(V)|MINimum|MAXimum
[:SOURce]:VOLTage[0]:LIMit:LOW?
0 ~ the present setting voltage
Minimum allowed voltage level
Maximum allowed voltage level
SOUR:VOLT:LIM:LOW MAX
Sets the UV> level to its maximum.