Gw-instek GDS-2000A series Programming Manual Bedienungsanleitung Seite 140

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 272
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 139
GDS-2000A Programming Manual
140
Return Parameter
Returns the triggering condition.
Example
:TRIGger:BUS:B1:UART:CONDition TXDATA
Sets the UART bus to trigger on Tx Data.
:TRIGger:BUS:B1:UART:RX:DATa:SIZe
Description
Sets or queries the number of bytes for UART data.
Note
This setting only applies when the UART trigger is
set to trigger on Rx Data
Syntax
:TRIGger:BUS:B1:UART:RX:DATa:SIZe {<NR1> | ?}
Related
commands
:TRIGger:BUS:B1:UART:CONDition
Parameter
<NR1>
Number of bytes (1 to 10).
Return parameter
<NR1>
Returns the number of bytes.
Example
:TRIGger:BUS:B1:UART:RX:DATa:SIZe 5
Sets the number of bytes to 5.
:TRIGger:BUS:B1:UART:RX:DATa:VALue
Description
Sets or queries the triggering data value for the
UART bus when the bus is set to trigger on Rx
Data.
Syntax
:TRIGger:BUS:B1:UART:RX:DATa:VALue {string | ? }
Related
commands
:TRIGger:BUS:B1:UART:RX:DATa:SIZe
Parameter
<sting>
The number of characters in the string
depends on the data size setting. The
string must be enclosed in double
quotes, "string".
x = don’t care
1 = binary 1
0 = binary 0
Set
Query
Set
Query
Seitenansicht 139
1 2 ... 135 136 137 138 139 140 141 142 143 144 145 ... 271 272

Kommentare zu diesen Handbüchern

Keine Kommentare