COMMAND DETAILS
57
Query Example
:CURR:STAT:REC?
0
A Value is active.
:CURRent:STATic:L1/L2
Channel Specific
Command
Description
Sets the A/B Value for constant current static
mode, where L1 is A Value and L2 is B Value. The
command is range dependant. If the current range
is Low, then the command will only apply to the
low range settings.
Syntax
:CURRent:STATic:L1|L2 <NRf+>[A]
Parameter
<NRf+>[A]
L1 1
Sets A Value to 1 Amp.
L2 2
Sets B Value to 2 Amps.
L1 1A
Sets A Value to 1 Amp.
(single channel only)
L1 MIN
Sets A Value to the
minimum level for the
specific channel.
L1 MAX
Sets A Value to the
maximum Level for the
specific channel.
Example
:CURR:STAT:L1 1
Sets A Value to 1 amp
for the current range
Query Syntax
:CURRent:STATic:L1?/L2? [MAX|MIN]
Return Parameter
<NR2> [MAX|MIN]
Current
1 unit = 1 amp
Returns the current of the
A Value (L1) or B Value
(L2).
MAX/MIN
Returns the allowable
maximum and minimum.
Query Example
:CURR:STAT:L2? MAX
10.2
Returns the maximum
current allowed for the
channel. (PEL-2020)