Gw-instek PEL-2000 Series Programming User Manual Bedienungsanleitung Seite 67

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 170
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 66
COMMAND DETAILS
67
:CURRent:DYNamic:LOW:T1/T2
Channel Specific
Command
Description
Sets the low range timers T1 or T2 for CC dynamic
mode for the specific channel.
Syntax
:CURRent:DYNamic:LOW:T1/T2 <NRf+>[S/ms]
Parameter
<NRf+>[S/ms]
Time
T1/T2 0.1S
Sets the T1/T2 time to 0.1
seconds.
T1/T2 1
Sets T1/T2 to 1 second.
T1/T2 MIN
Sets the T1/T2 to the
minimum value.
T1/T2 MAX
Sets the T1/T2 time to the
maximum time
Example
:CURR:DYNA:LOW:T1 .1S
Sets the T1 time to 100
milliseconds for the
specific channel.
Query Syntax
: CURRent:DYNamic:LOW:T1/T2? [MIN|MAX]
Return Parameter
<NR2> [MAX|MIN]
Time
1 Unit=1 second
Returns T1/T2 time.
MAX/MIN
Returns the allowable
maximum and minimum.
Query Example
:CURR:DYN:LOW:T1?
2.5
:CURR:DYN:LOW:T1?
MIN
0.000025
Returns the T1 time of
2.5 seconds.
Returns the minimum T1
time allowable for the
specific channel.
:CURRent:DYNamic:HIGH:L1/L2
Channel Specific
Command
Description
Sets the high range current levels (Level 1 & 2) for
CC dynamic mode.
Seitenansicht 66
1 2 ... 62 63 64 65 66 67 68 69 70 71 72 ... 169 170

Kommentare zu diesen Handbüchern

Keine Kommentare