PEL-3000 Programming Manual
82
Query example
:RESistance:L1?
> 0
Returns the “level1” conductance setting(MHO).
:RESistance:L2
Set
Query
Description
Sets or queries the CR mode “level2conductance.
This command is applicable to dynamic mode
only.
Note: A different value can be set for each current
range (High/Mid/Low).
Syntax
:RESistance:L2 {<NRf>[ MHO ] | MINimum |
MAXimum }
Query Syntax
:RESistance:L2?
Parameter
<NRf>
level2” conductance
value
MINIMUM or MIN
Minimum conductance
level
MAXIMUM or MAX
Maximum conductance
level
Return parameter
<NR2>
Returns the “Level2
conductance value.
Example
:RESistance:L2 MIN
Sets the conductance value to the minimum.
Query example
:RESistance:L2?
> 2.
Returns the “Level2conductance setting.(MHO)
:RESistance:SET
Set
Query
Description
Sets or queries the CR dynamic mode Set value
when the dynamic units are set to Percent.
This command is only applicable in dynamic
mode.