REMOTE CONTROL
113
:MMEMory:LOAD:CORRection
Set
Description
Loads correction data from a file to the internal
memory.
Syntax
:MMEMory:LOAD:CORRection <corr num>,
<src_file_name>
Parameter
<corr num>
<NR1> correction set 1~5
<src_file_name>
XXX.cor
Example
:MMEM:LOAD:CORR 2,test.cor
:MMEMory:LOAD:LIMit
Set
Description
Loads limit line data from a file to the internal
memory.
Syntax
:MMEMory:LOAD:LIMit <lim num>,<src_file_name>
Parameter
<lim num>
<NR1> limit line 1~5
<src_file_name>
XXX.lim
Example
:MMEM:LOAD:LIM 2,test.lim
:MMEMory:LOAD:PMETer
Set
Description
Loads power meter data from a file to the internal
memory.
Syntax
:MMEMory:LOAD:PMETer <src_file_name>
Parameter
<src_file_name>
XXX.pmet
Example
:MMEM:LOAD:PMET test.pmet
:MMEMory:LOAD:SEQuence
Set
Description
Loads sequence data from a file to the internal
memory.
Seitenansicht 114