[:SENSe]:SEMask:CARRier:AUTO {OFF|ON|0|1}
[:SENSe]:SEMask:CARRier:AUTO?
0
Turn PSDRef/TotalPwrRef to manual
(off).
1
Turn PSDRef/TotalPwrRef to auto (on).
OFF
Turn PSDRef/TotalPwrRef to manual
(off).
ON
Turn PSDRef/TotalPwrRef to auto (on).
0
PSDRef/TotalPwrRef is set to manual.
1
PSDRef/TotalPwrRef is set to automatic.
:SEM:CARR:AUTO OFF
[:SENSe]:SEMask:CARRier:CPSD
Set
Query
Sets or queries the value of the PSDRef for SEM
measurements.
[:SENSe]:SEMask:CARRier:CPSD <NR3>
[:SENSe]:SEMask:CARRier:CPSD?
<NR3>
PSD ref unit = dBm/Hz
:SEM:CARR:CPSD 20
[:SENSe]:SEMask:CARRier:POWer
Set
Query
Sets or queries the value of the TotalPwrRef
amplitude for SEM measurements.
[:SENSe]:SEMask:CARRier: POWer <ampl>
[:SENSe]:SEMask:CARRier: POWer?
<ampl>
<NRf>
<NR3>
dBm
:SEM:CARR:POW 2 dbm